Technical Specifications
Parameter | SuperLabED 20 | SuperLabED 40 |
---|---|---|
Generation performance @ 15°C | 20 | 40 |
Conductivity µS/cm | < 0.1 | < 0.1 |
Resistivity MΩ·cm | > 10 | > 10 |
TOC ppb | < 30 | < 30 |
Dispensing Rate l/min | Upto 2 | Upto 2 |
Pure water Pressurized outlet | Standard | Standard |
Retention of Bacteria, Particles and Pyrogens by RO Stage | 95 – 99% | 95 – 99% |
Particles 0.2 µm/ml | No particles ≥ 0.2µm | No particles ≥ 0.2µm |
Bacterial Content CFU/ml | < 0.01 | < 0.01 |
UV Photo-Oxidation (254nm) | Optional | Optional |
General Technical Data | ||
Operating Pressure bar | 2 – 6 | 2 – 6 |
Supply Voltage Volts/Hz | 90 – 240/50–60 | 90 – 240/50–60 |
Power Consumption kW | 0.1 | 0.1 |
Water Connections | 8mm | 8mm |
Ambient Temperature | 2 – 35 | 2 – 35 |
Dimensions W x H x D mm | 511 x 1520 x 575 | 511 x 1520 x 575 |
Weight | 41 | 41 |
Feedwater Requirements | ||
Conductivity | < 2500 | < 2500 |
TOC | < 2000 | < 2000 |
Chlorine Content | < 5 | < 5 |
Silt Density / Fouling Index (SDI or FI) | < 12 | < 12 |
Manganese and Iron content mg/l | < 0.1 | < 0.1 |
Temperature °C | 2 – 35 | 2 – 35 |
Hardness | Upto 300 ppm | Upto 300 ppm |